OxyTrace Oxy-333-XL Electrochemical Galvanic Trace O2 Sensor for PPM Oxygen Measurement (zero CO2 cross sensitivity)
Oxy-333-XL PPM Oxygen Sensor - Replacement to Analytical Industries Model XLT-12-333 and OXY-T-2 PPM oxygen sensor. Oxy--333-XL PPM O2 sensor has a significantly higher signal output per PPM thus giving great signal to noise ratio with high sensitivity. Its proprietary electrolyte formulation offers excellent compatibility of the sensor with sample gas containing high levels of CO2 and trace levels of H2S and SO2 gases. All Oxy-333-XL PPM O2 sensors are gas tested for output, stability and recovery from air to low PPM levels before shipment. Sensor is designed, developed and manufactured in the China. Tags: OXY-T-2, XLT-12-333
0-10,000 ppm range
Signal output(10,000ppm)600μA±120μA ambient air
Electrical connections-2ring PCB
Key features Measuring range 0-10,000 ppm O2 Signal output (10,000ppm)-600μA±120μA ambient air Response time (T90) - 3 seconds Operating temperature - 0-50°C Operating pressure range - 0.5-2bar absolute Expected service life - 24 months Electrical connections - 2 ring PCB Flow rate 0-2.5 LPM ChangAiModel Oxy-333-XL PPM O2 Sensor Replaces Analytical Industries Model XLT-12-333 and XLT-T-2
Overview
This sensor is specifically designed for oxygen measurement in CO2 background gases. This innovative sensor design carries over many of the features of the traditional galvanic oxygen sensor but has a much higher sensitivity, stability, and signal-to-noise ratio with a 24-month operating life that offers an effective sensing solution for measuring low levels of PPM oxygen